MeaTest 9010+ Multifunction Calibrator
10 ppm AC/DC multifunction calibrator for multimeters, power meters and generally all kinds of electric meters. Implementing modular design, 9010+ can be further customized to calibrate oscilloscopes, insulation testers, sources and transducers.
- 10 ppm accuracy
- 400 MHz and 1.1 GHz scope options
- Power quality functions, voltage from current
Multifunction calibrator 9010+ is designed as universal calibration tool for electrical calibration laboratories, covering most of their workload like 6½ digit multimeters, clamp meters, ohm meters, power meters and power analysers, energy meters, transducers, insulation testers, process meters, oscilloscopes and many others. High load capacity of voltage output (up to 50 mA) allows for calibration of high-consumption analogue meters. Installed harmonic and non-harmonic shape signals allow for testing meter sensitivity to distorted signals by a signal with various crest factor.9010+ takes advantage of 9010’s wide workload coverage and modern user interface ans pushes its performance even further by accuracy improvements in most functions, current range extension to 30 A, power range extension to 1575 kW and ultimately 1.1 GHz o-scope calibration capability with SC1 option.
DC Voltage |
0 – 1050 V |
---|---|
AC Voltage |
1 mV – 1050 V, 15 Hz – 300 kHz |
DC Current |
0 – 30 A (1500 A with 140-50 Current Coil) |
AC Current |
10 uA – 30 A, 15 Hz – 10 kHz (1500 A with 140-50 Current Coil) |
DC Power & Energy |
40 uW – 30 kW (1575 kW with 140-50 Current Coil) |
AC Power & Energy |
40 uW – 30 kW, 15 – 1000 Hz (1575 kW with 140-50 Current Coil) |
Resistance |
0 Ω – 1.1 GΩ, resolution 6½ digits |
Insulation Resistance |
100 kΩ – 100 GΩ |
Capacitance |
800 pF – 120 mF, resolution 6½ digits |
Temperature |
TC BCDEG2JKMNRST with cold junction compensation |
Frequency |
0.1 Hz – 400 MHz |
MeaTest 9010+ Multifunction Calibrator
10 ppm AC/DC multifunction calibrator for multimeters, power meters and generally all kinds of electric meters. Implementing modular design, 9010+ can be further customized to calibrate oscilloscopes, insulation testers, sources and transducers.
- 10 ppm accuracy
- 400 MHz and 1.1 GHz scope options
- Power quality functions, voltage from current
Multifunction calibrator 9010+ is designed as universal calibration tool for electrical calibration laboratories, covering most of their workload like 6½ digit multimeters, clamp meters, ohm meters, power meters and power analysers, energy meters, transducers, insulation testers, process meters, oscilloscopes and many others. High load capacity of voltage output (up to 50 mA) allows for calibration of high-consumption analogue meters. Installed harmonic and non-harmonic shape signals allow for testing meter sensitivity to distorted signals by a signal with various crest factor.9010+ takes advantage of 9010’s wide workload coverage and modern user interface ans pushes its performance even further by accuracy improvements in most functions, current range extension to 30 A, power range extension to 1575 kW and ultimately 1.1 GHz o-scope calibration capability with SC1 option.
DC Voltage |
0 – 1050 V |
---|---|
AC Voltage |
1 mV – 1050 V, 15 Hz – 300 kHz |
DC Current |
0 – 30 A (1500 A with 140-50 Current Coil) |
AC Current |
10 uA – 30 A, 15 Hz – 10 kHz (1500 A with 140-50 Current Coil) |
DC Power & Energy |
40 uW – 30 kW (1575 kW with 140-50 Current Coil) |
AC Power & Energy |
40 uW – 30 kW, 15 – 1000 Hz (1575 kW with 140-50 Current Coil) |
Resistance |
0 Ω – 1.1 GΩ, resolution 6½ digits |
Insulation Resistance |
100 kΩ – 100 GΩ |
Capacitance |
800 pF – 120 mF, resolution 6½ digits |
Temperature |
TC BCDEG2JKMNRST with cold junction compensation |
Frequency |
0.1 Hz – 400 MHz |